-
Products
- Gas analysis systems
- GAOS SENSON gas analyzers
- GAOS MS process mass spectrometry
- MaOS HiSpec ion mobility spectrometer
- MaOS AxiSpec ion mobility spectrometer
- Applications
- News
- Events
- About us
RAMOS E200 | RAMOS M350 | RAMOS M520 | RAMOS M750 | ||||
Image acquisition mode: |
3D (XYZ) confocal laser and Raman images |
||||||
Scanning type: |
XY galvano-mirror scanners XY motorized stage (option) Z piezo scanner |
||||||
Scanning speed: |
1000 х 1000 pixels per 3 sec (3 μs/pixel) |
||||||
Spatial resolution: | XY: 440 nm, Z: 620 nm (532 nm laser, 100 x , NA = 0.95) | ||||||
Spectral range: |
50 – 8500 cm-1 (532 nm laser) |
50 – 9700 cm-1 (532 nm laser) |
|||||
Excitation source: |
Build-in 473 nm or 532 nm laser, connection of additional external lasers 455 nm, 633 nm, 785 nm is possible |
||||||
Laser beam attenuator: |
Automated unit with VND filter, continuously changeable from 0.1% to 100% |
||||||
Rayleigh rejection filters: |
A pair of Edge filters with cut-off starting from 50 cm-1 (for 532 nm laser) |
||||||
Spectrometer configuration: |
2-channel imaging spectrometer directly coupled to a microscope |
External imaging spectrograph |
|||||
Focal length: | 200 mm | 350 mm | 520 mm | 750 mm | |||
Spectral resolution |
1 cm-1/pixel (grating 2400 gr/mm) 2 cm-1/pixel, FWHM (grating 2400 gr/mm) |
1.60 cm-1/pixel (grating 1800 groove/mm) |
0.76 cm -1/pixel (grating 1800 groove/mm) 0.25 cm -1/pixel (Echelle grating) |
0.44 cm -1/pixel (grating 1800 groove/mm) |
|||
Number of gratings: |
2 gratings, 4 optionally |
4 mounted on motorized turret, more gratings with manual changing |
|||||
Detection system: |
CCD sensor 2048х122 pixels, with Peltier cooling |
CCD sensor 2048х122 with two-stage cooling, quantum efficiency up to 95% |
|||||
Options: |
Heating & cooling stages, fiber optic probe, |